The Grant Optima™ GP200 series of digitally controlled stirred water baths and circulators provides outstanding performance for routine and sophisticated applications requiring accurate temperature control up to a maximum of 200ºC.
Grant GP200 series systems combine the powerful and versatile Grant Optima™ GP200 digital immersion thermostat with a choice of three economy plastic (P) tanks or five stainless steel (S) tanks with different capacities.
The Grant GP200 immersion thermostat incorporates Intelligent Control Optimisation™ (ICO) for adaptive intelligent PID temperature control and includes a powerful integral pump, making GP200 systems suitable both for immersing samples and circulating temperature controlled fluid to external devices. The temperature range and heater power are automatically limited according to the liquid type selected.
The GP200 control unit has a host of powerful and convenient programming features for optimum application flexibility. These include a convenient heater timer for early starts/late shutdowns, two programmable relays for switching external devices on/off, option for remote temperature monitoring, RS232 interface for PC communication and data collection, and memory capacity for five programs of up to 30 segments each.
- Dual display: bright LED for temperature display plus 2-line 16 character LCD for easy programming
- Twin heaters, 2kW heater power for fast heat up
- Pt1000 external probe socket for remote monitoring
- Powerful integral pump with multi-stage variable flow rate, maximum flow 21L/min
- Variable high and low temperature alarm settings - configurable to switch relays
- Choice of 5 operator languages
- Optional Labwise™ PC software for program set-up, data logging and real-time graphing
- Maximum cooling/heating range (dependent on tank size): ambient +5 to 99ºC with plastic (P) tank, or -15 to 200ºC with stainless steel (S) tank
- Stability ± 0.005ºC
Note: Alternatively, the Grant Optima™ GP200 immersion thermostat can also be used to regulate the temperature in vessels other than Grant tanks. Such vessels must have vertical sides and a maximum wall thickness of 35 mm for rectangular tanks, 30 mm for circular tanks (300 mm diameter) and a capacity of up to 50 litres. The control unit is attached to other vessels by means of a clamp.
The powerful Grant Optima™ GP200 series of water baths and circulators with plastic (P) tanks represents excellent performance and value for money for more demanding applications requiring accurate temperature control in the range of ambient plus 5 to 99ºC.
- Robust plastic construction, double-walled for rigidity, easy to clean; designed for use with water
- Convenient recessed handholds for carrying/repositioning the unit
- Heating range ambient +5 to 99ºC (ambient +15 to 99ºC for P5 5-litre tank)
- Stability ± 0.005ºC
- Choice of three different plastic tanks sizes:
Bath model GP200-P5 - 5 litre
Bath model GP200-P12 - 12 litre
Bath model GP200-P18 - 18 litre
Accessory options for GP200-P series baths
Lids and polypropylene spheres
Between 60ºC and 99ºC a lid or layers of polypropylene spheres should be used to help reduce evaporation/heat loss and avoid sample contamination. Polypropylene spheres are particularly useful where tall vessels need to be immersed.
Lids
PL5 flat stainless steel lid for GP200-P5
PL12 curved plastic lid for GP200-P12
PL18 curved plastic lid for GP200-P18
Polypropylene spheres
PS20 pack of polypropylene spheres:
1 pack required for GP200-P5 and for GP200-P12
2 packs required for GP200-P18
Rack systems and raised shelves
Grant custom designed racks and shelves enable you to optimise the available baths capacity. Raised shelves allow shallow vessels to be accommodated.
Racks
QR rack for GP200-P5 - 1 QR rack per bath
| Tube size |
Number of tubes per QR |
| Ø 10-13 mm |
30 |
| Ø 16-19 mm |
16 |
| Ø 24 mm |
10 |
| Ø 30 mm |
5 |
| 0.5 ml |
44 |
| 1.5 ml |
35 |
VR rack for GP200-P12: - 2 VR racks per bath
VR rack for GP200-P18: - 4 VR racks per bath
| Tube size |
Number of tubes per VR |
| Ø 10-13 mm |
65 |
| Ø 16-19 mm |
36 |
| Ø 24 mm |
23 |
| Ø 30 mm |
14 |
| 0.5 ml |
102 |
| 1.5 ml |
75 |
Raised shelves
Reduce the effective liquid depth to allow shallow vessels to be placed in the baths. Not available for GP200-P5.
RS14 raised shelf for GP200-P12
RS22 raised shelf for GP200-P18
Accessory vertical turbine pumps
Required only where the application demands a higher pressure than that delivered by the GP200 internal pump to maintain flow. Note: When an accessory pump is fitted, the available working area is reduced.
Low noise, compact design. Supplied with pipe connections and special lid for fitting to tank, pipe bore 12.7 mm.
VTP1 - max. pressure 1000 mbar, max. flow 9 L/min
VTP2 - max. pressure 1650 mbar, max. flow 12 L/min
External probes
for monitoring and controlling temperature of remote loads
FF17 flexible polythene probe - 2 m cable, 100mm x Ø 4.5mm, very fast response
LL17 stainless steel probe - 2 m cable, 125mm x Ø 5mm
Remote switching device
RSD remote switching device
Can be used by GP200 models to control, by switching on and off, any mains powered appliance up to a maximum of 8 Amps. The GP200 can control two relays. A separate RDS is required for each relay. This function can be programmed directly via the GP200 interface or via PC with Labwise™ software.
Labwise™ PC software
Allows two-way communication for status display, programming and data capture
The powerful Grant Optima™ GP200 series of water baths and circulators with stainless steel (S) tanks provides outstanding performance for more demanding heating and cooling applications requiring accurate temperature control up to 200ºC.
- Robust plastic outer case and stainless steel tank - easy to clean, resists corrosion
- Dual position fitting of the thermostat ensures convenient access in all situations
- Convenient recessed handholds for carrying/repositioning the unit
- Maximum heating/cooling* range -15 to 200ºC (dependent on tank size - see below)
- Stability ± 0.005ºC
- Choice of five different stainless steel tanks sizes:
Bath model GP200-S5 - 5 litre (0 to 200ºC)
Bath model GP200-S12 - 12 litre (0 to 200ºC)
Bath model GP200-S18 - 18 litre (0 to 200ºC)
Bath model GP200-S26 - 26 litre (-15 to 200ºC)
Bath model GP200-S38 - 38 litre (-15 to 200ºC)
*Minimum operating temperature without accessory cooling is room temperature plus 5ºC (room temperature plus 15ºC for S5 tanks).
Accessory options for GP200-S series baths
Lids and polypropylene spheres
Between 60ºC and 100ºC and below room temperature a lid or layers of polypropylene spheres should be used to help reduce evaporation/heat loss and avoid sample contamination. Above 100ºC a lid must be used.
Note: Polypropylene spheres are particularly useful where tall vessels need to be immersed.
Lids
FG5 flat stainless steel lid for GP200-S5
LG12 gabled, hinged, removable stainless steel lid for GP200-S12
LG18 gabled, hinged, removable stainless steel lid for GP200-S18
LG26 gabled, hinged, removable stainless steel lid for GP200-S26
LG38 gabled, hinged, removable stainless steel lid for GP200-S38
Polypropylene spheres
PS20 pack of polypropylene spheres:
1 pack required for GP200-S5 and GP200-S12
2 packs required for GP200-S18 and GP200-S26
3 packs required for GP200-S38
Rack systems and raised shelves
Grant custom designed racks and shelves enable you to optimise the available baths capacity. Raised shelves allow shallow vessels to be accommodated.
Racks
QR rack for GP200-S5 - 1 QR rack per bath
| Tube size |
Number of tubes per QR |
| Ø 10-13 mm |
30 |
| Ø 16-19 mm |
16 |
| Ø 24 mm |
10 |
| Ø 30 mm |
5 |
| 0.5 ml |
44 |
| 1.5 ml |
35 |
VR rack for
GP200-S12: -
2 VR racks per bath
VR rack for
GP200-S18: -
4 VR racks per bath
VR rack for
GP200-S26: -
4 VR racks per bath
VR rack for
GP200-S38: -
6 VR racks per bath
| Tube size |
Number of tubes per VR |
| Ø 10-13 mm |
65 |
| Ø 16-19 mm |
36 |
| Ø 24 mm |
23 |
| Ø 30 mm |
14 |
| 0.5 ml |
102 |
| 1.5 ml |
75 |
Raised shelves
Reduce the effective liquid depth to allow shallow vessels to be placed in the baths. Not available for GP200-S5.
RS14 raised shelf for GP200-S12
RS22 raised shelf for GP200-S18
RS28 raised shelf for GP200-S26
RS28 or RS38 raised shelf for GP200-S38
Accessory cooling systems
To allow systems to operate at or below room temperature by means of a cooling coil dipped into the bath. Designed for minimal impact on working area.
Refrigerated immersion coolers
Consist of a cooling coil connected to a refrigeration unit by a flexible pipe. Extract heat continuously, with the bath control unit controlling temperature.
C1G immersion cooler , temperature range 0 to 40ºC, suitable for all GP200-S series baths
C2G immersion cooler, temperature range -15 to 40ºC, suitable for GP200-S26 and GP200-S38
Heat exchange coil
Designed to be attached to a supply of cooling tap water or a refrigerated circulator
CW5 heat exchange coil , temperature range 2ºC above coolant temperature, suitable for all GP200-S series baths
Accessory vertical turbine pumps
Required only where the application demands a higher pressure than that delivered by the GP200 internal pump to maintain flow. Note: When an accessory pump is fitted, the available working area is reduced.
Low noise, compact design. Supplied with pipe connections and special lid for fitting to tank, pipe bore 12.7 mm.
VTP1 - max. pressure 1000 mbar, max. flow 9 L/min
VTP2 - max. pressure 1650 mbar, max. flow 12 L/min
External probes
for monitoring and controlling temperature of remote loads
FF17 flexible polythene probe, 2 m cable, 100mm x Ø 4.5mm, very fast response
LL17 stainless steel probe, 2 m cable, 125mm x Ø 5mm
Remote switching device
RSD remote switching device
Can be used by GP200 models to control, by switching on and off, any mains powered appliance up to a maximum of 8 Amps. The GP200 can control two relays. A separate RDS is required for each relay. This function can be programmed directly via the GP200 interface or via PC with Labwise™ software.
Labwise™ PC software
Allows two-way communication for status display, programming and data capture
| |
|
|
GP200-S5 |
GP200-S12 |
GP200-S18 |
GP200-S26 |
GP200-S38 |
| Temperature range 1 |
|
ºC |
0 to 200 2 |
0 to 200 3 |
0 to 200 3 |
-15 to 200 3 |
-15 to 200 3 |
| Stability (Din 58966) |
@37°C |
±ºC |
0.005 |
| Uniformity |
@37°C |
±ºC |
0.02 |
| Setting resolution |
|
ºC |
0.1 |
| |
|
|
|
| Display |
|
|
4 digit 13mm LED |
| |
|
|
2 line 16 character LCD |
| Display resolution |
|
ºC |
0.01 (LCD) |
| No. stored temperature values |
|
|
4 |
| Two point re-calibration |
|
|
yes |
| Offset adjustment |
|
|
yes |
| External probe |
|
|
Pt1000 |
| RS232 |
|
|
yes |
| Programmable |
|
|
remote via PC / direct |
| No. stored programs |
|
|
5 x 30 segment |
| Relays |
|
|
2 |
| Heater power |
240V |
kW |
2 |
| |
115V |
kW |
1.3 |
| Pump |
max. pressure |
mbar |
460 |
| (water) |
max. flow |
L/min |
21, adjustable flow rate |
| Pipe bore |
inlet/outlet |
mm |
6, 11 |
| Safety |
overtemp. |
|
adjustable cut-out |
| |
level |
|
float switch |
| Alarms |
|
|
high and low |
| Electrical power |
220-240V |
kW |
2.2 (50Hz) |
| |
110-120V |
kW |
1.4 (60Hz) |
| |
|
|
|
|
|
|
|
| Tank capacity |
|
L |
5 |
12 |
18 |
26 |
38 |
| Top opening |
l/w |
mm |
150/150 |
210/300 |
390/300 |
390/300 |
580/300 |
| Liquid depth |
min/max. |
mm |
80/140 |
80/140 |
70/130 |
120/180 |
120/180 |
| Inner tank dimensions |
l/w/h |
mm |
300/150/150 |
325/300/150 |
505/300/150 |
505/300/200 |
690/300/200 |
| Outer case dimensions |
l/w/h |
mm |
325/175/175 |
350/325/175 |
530/325/225 |
530/325/225 |
720/325/225 |
| Overall dims. inc. controller |
l/w/h |
mm |
325/175/355 |
350/325/355 |
530/325/405 |
530/325/405 |
720/325/405 |
| |
|
|
|
|
|
|
|
| Suitable immersion cooler |
|
|
C1G |
C1G |
C1G |
C1G, C2G |
C1G, C2G |
| |
|
|
GP200-P5 |
GP200-P12 |
GP200-P18 |
| Temperature range |
|
ºC |
amb + 5 to 99 |
| Stability (Din 58966) |
@37ºC |
±ºC |
0.005 |
| Uniformity |
@37ºC |
±ºC |
0.02 |
| Setting resolution |
|
ºC |
0.1 |
| |
|
|
|
| Display |
|
|
4 digit 13mm LED |
| |
|
|
2 line 16 character LCD |
| Display resolution |
|
ºC |
0.01 (LCD) |
| No. stored temperature values |
|
|
4 |
| Two point re-calibration |
|
|
yes |
| Offset adjustment |
|
|
yes |
| External probe |
|
|
Pt1000 |
| RS232 |
|
|
yes |
| Programmable |
|
|
remote via PC / direct |
| No. stored programs |
|
|
5 x 30 segment |
| Relays |
|
|
2 |
| Heater power |
240V |
kW |
2 |
| |
115V |
kW |
1.3 |
| Pump |
max. pressure |
mbar |
460 |
| (water) |
max. flow |
L/min |
21, adjustable flow rate |
| Pipe bore |
inlet/outlet |
mm |
6, 11 |
| Safety |
overtemp. |
|
adjustable cut-out |
| |
level |
|
float switch |
| Alarms |
|
|
high and low |
| Electrical power |
220-240V |
kW |
2.2 (50Hz) |
| |
110-120V |
kW |
1.4 (60Hz) |
| |
|
|
|
|
|
| Tank capacity |
|
L |
5 |
12 |
18 |
| Top opening |
l/w |
mm |
150/120 |
303/220 |
400/303 |
| Liquid depth |
min./max. |
mm |
80/140 |
80/140 |
80/140 |
| Inner tank dimensions |
l/w/h |
mm |
240/160/150 |
335/285/150 |
515/290/150 |
| Outer case dimensions |
l/w/h |
mm |
330/240/180 |
420/350/180 |
605/370/180 |
| Overall dimensions inc. controller |
l/w/h |
mm |
330/240/360 |
420/350/360 |
605/370/360 |
1 A lid or layer of polypropylene spheres must be used when operating between 60ºC and 100ºC, and is recommended for use below room temperature.
2 Minimum operating temperature without accessory cooling is room temperature plus 15ºC